18 MG色譜柱(250 mm×4.6 mm,5μm);流動相為0.1%磷酸水溶液-乙腈(97.5:2.5);檢測波長為215 nm,體積流量為1 mL/min;柱溫為35℃;進樣量為20 μL。結(jié)果 乳酸在0.1~5.0 mg/mL與峰面積呈良好的線性關(guān)系(r=0.999 9),平均回收率為99.8%,RSD值為0.8%(n=9)。結(jié)論 此方法操作便捷、經(jīng)濟,方法準確度、精密度、耐用性好,可作為藥用輔料乳酸中乳酸測定的分析方法。;Objective To establish an HPLC method for assay of lactic acid in pharmaceutical excipient lactic acid. Methods HPLC method was adopted. The determination was carried out on Capcell Pak C18 MG C18 column (250 mm×4.6 mm, 5 μm). The mobile phase was consisted of 0.1% phosphoric acid aqueous solution-acetonitrile (97.5:2.5). The detection wavelength was set at 215 nm. The flow rate was 1.0 mL/min. The column temperature was 35℃, and volume of injection was 20 μL. Results The linear relation was good (r=0.999 9) between lactic acid and peak area in the range of 0.1-5.0 mg/mL. The average recovery rate was 99.8% with RSD values of 0.8% (n=9). Conclusion The method has simple operation, good accuracy, precision and durability, which can be performed in assay of lactic acid in pharmaceutical excipient lactic acid."/>